Remote control commands
R&S
®
SMA100B
455User Manual 1178.3834.02 ─ 09
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
*OPC).
Usage: Event
14.4 Preset commands
The preset commands are not bundled in one subsystem. Therefore, they are listed
separately in this section.
Four presetting actions are available:
Activating the default state of all internal instrument functions (*RST on page 453).
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. reference oscillator source settings.
Activating the preset state of the parameters related to the selected signal path (:
SOURce<hw>:PRESet on page 455)
Activating the preset state of all parameters that are not related to the signal path
(:DEVice:PRESet on page 455)
Activating the original state of delivery (factory reset, :SYSTem:FPReset
on page 456). Only functions that are protected by a password remain unchanged
as well as the passwords themselves.
When resetting, the following deviation between remote and manual control exists:
In contrast to the [Preset] key, the SCPI commands *RST and :SYSTem:PRESet do
not close open dialogs in the GUI.
:DEVice:PRESet............................................................................................................ 455
:SOURce<hw>:PRESet.................................................................................................. 455
:SYSTem:PRESet.......................................................................................................... 456
:SYSTem:FPReset......................................................................................................... 456
:DEVice:PRESet
Presets all parameters which are not related to the signal path, including the LF gener-
ator.
Example:
DEV:PRES
Presets all instruments settings that are not related to the signal
path.
Usage: Event
:SOURce<hw>:PRESet
Presets all parameters which are related to the selected signal path.
Preset commands