List and sweep mode
R&S
®
SMA100B
189User Manual 1178.3834.02 ─ 09
Remote command:
[:SOURce<hw>]:FREQuency:MODE on page 643
State (RF level sweep)
Activates RF level sweep signal generation.
Note: Active RF frequency, RF level or RF combined sweep modes deactivate the LF
sweep or List mode and vice versa.
Remote command:
[:SOURce<hw>]:POWer:MODE on page 687
State (RF frequency / level sweep)
Activates the RF frequency / level sweep signal generation.
Note: Active RF frequency, RF level or RF combined sweep modes deactivate the LF
sweep or List mode and vice versa.
Remote command:
[:SOURce<hw>]:FREQuency:MODE on page 643
[:SOURce<hw>]:POWer:MODE on page 687
State (LF frequency sweep)
Status of the LF frequency sweep signal.
You find this parameter in two dialogs:
●
"LF Frequency Sweep" turns on/off the signal generation.
See Sweep mode settings for access.
●
"Analog Modulation Sources" displays the current state (read-only). If it is turned
on, this dialog provides the varying sweep frequency, see Current Frequency.
Note: Active RF frequency, RF level or RF combined sweep modes deactivate the LF
sweep or List mode and vice versa.
Remote command:
[:SOURce<hw>]:LFOutput:FREQuency:MODE on page 656
Current Frequency
Displays the current frequency sweep value for all sweep Modes, except of Mode >
Step. In this mode, you can enter the next frequency setting manually.
Remote command:
[:SOURce<hw>]:FREQuency:MANual on page 645
[:SOURce<hw>]:LFOutput:FREQuency:MANual on page 655
Current Level
Displays the current level sweep value for all sweep Modes, except of Mode > Step. In
this mode, you can enter the next level setting manually.
Remote command:
[:SOURce<hw>]:POWer:MANual on page 686
Mode
Selects the sweep mode.
Sweep mode settings