EasyManua.ls Logo

R&S SMA100B - Page 656

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMA100B
656User Manual 1178.3834.02 ─ 09
Parameters:
<Manual> float
You can select any value within the setting range, where:
STARt is set with [:SOURce<hw>]:LFOutput:FREQuency:
STARt
STOP is set with [:SOURce<hw>]:LFOutput:FREQuency:
STOP
Range: STARt to STOP
Increment: 0.1
*RST: 1000
Example: See Example "Setup an LF sweep" on page 652.
Manual operation: See "Current Frequency" on page 189
[:SOURce<hw>]:LFOutput:FREQuency:MODE
<Mode>
Sets the mode for the output of the LF generator frequency, and determines the com-
mands to be used for frequency settings.
Parameters:
<Mode> CW | FIXed | SWEep
CW|FIXed
Sets the fixed-frequency mode. CW and FIXed are synonyms.
To set the output frequency, use command [:SOURce]:
LFOutput<ch>:FREQuency
SWEep
Sets sweep mode.
To set the frequency, use the commands:
[:SOURce<hw>]:LFOutput:FREQuency:STARt and [:
SOURce<hw>]:LFOutput:FREQuency:STOP
Or [:SOURce<hw>]:LFOutput:FREQuency:MANual
*RST:
CW
Example: See Example "Setup an LF sweep" on page 652.
Manual operation: See "State (LF frequency sweep)" on page 114
[:SOURce<hw>]:LFOutput:FREQuency:STOP <Stop>
[:SOURce<hw>]:LFOutput:FREQuency:STARt <Start>
Sets the start/stop frequency for [:SOURce<hw>]:LFOutput:FREQuency:MODE
SWEep.
Parameters:
<Start> float
Range: 0.1 Hz to 1 MHz
Increment: 0.1
*RST: 1 KHz
SOURce subsystem

Table of Contents

Other manuals for R&S SMA100B

Related product manuals