Network operation and remote control
R&S
®
SMA100B
419User Manual 1178.3834.02 ─ 09
The dialog shows all specified resource strings of the supported remote control
interfaces.
Note: Using the RS232 serial interface via USB requires the USB serial adapter
R&S TS-USB1.
13.8.2 How to change the GPIB instrument address
Option: R&S SMAB-B86
To control the instrument remotely via the GPIB bus, it must be addressed using the
GPIB address. The remote control address is factory-set to 28, but it can be changed if
it does not fit in the network environment. For remote control, addresses 0 through 30
are allowed. The GPIB address is maintained after a reset of the instrument settings.
To set the GPIB address:
1. Select "System Config > Remote Access > GPIB Address".
2. Select "GPIB Channel Address" and enter a value between 0 and 30.
Risk of losing remote connection
If the remote access and network settings had been configured to values different to
the default, executing a factory preset via remote control terminates the connection to
the instrument.
Controlling the R&S
SMA100B remotely