Remote control commands
R&S
®
SMA100B
501User Manual 1178.3834.02 ─ 09
Options: R&S SMAB-K28
Manual operation: See "Pow (Y)" on page 254
14.14.2 DISPlay subsystem
:DISPlay[:WINDow][:POWer]:SWEep:BACKground:COLor.................................................501
:DISPlay[:WINDow][:POWer]:SWEep:GRID:STATe............................................................501
:DISPlay[:WINDow][:POWer]:SWEep:BACKground:COLor <Color>
Defines the background color of the measurement diagram. The selected color applies
also to the hardcopy of the diagram.
Parameters:
<Color> BLACk | WHITe
*RST: BLACk
Example:
DISP:SWE:BACK:COL WHIT
The measurement is indicated with a white background.
Manual operation: See "Background Color - Power Analysis" on page 264
:DISPlay[:WINDow][:POWer]:SWEep:GRID:STATe <State>
Indicates a grid in the diagram.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example:
DISP:SWE:GRID:STAT OFF
Deactivates the indication of a grid in the diagram area.
Manual operation: See "Show Grid - Power Analysis" on page 264
14.14.3 INITiate command
:INITiate<hw>[:POWer]:CONTinuous <Continuous>
Switches the local state of the continuous power measurement by R&S NRP power
sensors on and off. Switching off local state enhances the measurement performance
during remote control.
The remote measurement is triggered with :READ<ch>[:POWer]?). This command
also returns the measurement results. The local state is not affected, measurement
results can be retrieved with local state on or off.
Parameters:
<Continuous> 0 | 1 | OFF | ON
*RST: 0
Power sensor measurement subsystems