This function is useful, for example to execute an SCPI sequence automatically but
with a defined time delay between some commands.
See Chapter 12.2.4, "How to assign actions to the [ (User)] key", on page 353.
Setting parameters:
<TimeMs> integer
Wait time in ms
Range: 0 to 10000
*RST: 0
Example:
SYSTem:WAIT 10000
// Waits 10s before resetting the instrument.
*RST
Usage: Setting only
:SYSTem:FILes:TEMPorary:DELete
Deletes the temporary files from the internal memory or, if installed, from the SD card
slot.
Example:
:SYSTem:FILes:TEMPorary:DELete
// temporary files are removed
Usage: Event
14.18 STATus subsystem
This system contains the commands for the status reporting system. See also Chap-
ter 13.4, "Status reporting system", on page 388 for detailed information.
*RST on page 453 has no effect on the status registers.
Value ranges
Queries return the current value of the respective register, which permits a check of
the device status.
Return values: A decimal value in the range 0 to 32767 (=2
15
-1)
The configuration commands set the respective register thus determining which
status changes of the R&S SMA100B cause the status registers to be changed.
Setting values: A decimal value in the range 0 to 32767 (=2
15
-1)
:STATus:OPERation:CONDition?.....................................................................................738
:STATus:OPERation:ENABle...........................................................................................738
:STATus:OPERation[:EVENt]...........................................................................................738
:STATus:OPERation:NTRansition.................................................................................... 738
:STATus:OPERation:PTRansition.....................................................................................739
:STATus:PRESet............................................................................................................739
:STATus:QUEStionable:CONDition.................................................................................. 739
:STATus:QUEStionable:ENABle.......................................................................................739
STATus subsystem

Table of Contents

Other manuals for R&S SMA100B

Related product manuals