Remote control commands
R&S
®
SMA100B
469User Manual 1178.3834.02 ─ 09
Example:
CAL:ALL:TIME?
// "22 days"
Usage: Query only
Manual operation: See "Time" on page 763
:CALibration:DATA:FACTory:DATE?
Queries the date of the last factory calibration.
Return values:
<Date> string
Example:
CAL:DATA:FACT:DATE?
// "2016-01-01"
Usage: Query only
Manual operation: See "Last Factory Calibration" on page 753
:CALibration<hw>:CONTinueonerror <State>
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
Suffix:
<hw>
.
[1]
Optional suffix
Parameters:
<State> 1 | ON | 0 | OFF
*RST: n.a. (factory preset: 0)
Example:
CAL:CONT ON
// Continues calibration after an error
Manual operation: See "Continue Adjustment on Error" on page 763
14.7 CSYNthesis subsystem
This subsystem contains the commands used to define the frequency settings for the
separate clock source (clock synthesis).
Option: see "Required options" on page 304.
CSYNthesis subsystem