EasyManua.ls Logo

R&S SMA100B - Page 753

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting and notifications
R&S
®
SMA100B
753User Manual 1178.3834.02 ─ 09
The "Hardware Config" dialog lists all installed assemblies and externally connec-
ted instruments with information on their part and serial numbers, and revision
states. The BIOS version is also listed; firmware updates do not update the BIOS
version.
The dialog is divided in tabs, according to the hardware components of the signal
domains. The "Counter" tab provides information on the operation time and number
of times the instrument was powered on.
The remote commands required to query the hardware configuration are described in
Chapter 14.8, "DIAGnostic subsystem", on page 474.
Assembly
The tables in the tabs show characteristics of the installed assemblies.
"Assembly"
Assembly designation.
"Part Number"
Part number of the assembly.
"Serial Number"
Serial number of the assembly.
"Revision"
Revision state of the assembly.
"Slot"
Indicates whether the assembly is connected to the serial bus or PCI
bus.
Remote command:
:DIAGnostic<hw>:BGINfo? on page 475
Counter
Displays information on the operation times of the R&S SMA100B.
Operation Time / h ← Counter
Displays the operation time in hours so far.
Remote command:
:DIAGnostic:INFO:OTIMe? on page 476
Power On Count ← Counter
Displays the number the instrument has been turned on.
Remote command:
:DIAGnostic:INFO:POCount? on page 476
Last Factory Calibration ← Counter
Displays the date of the last factory calibration.
Remote command:
:CALibration:DATA:FACTory:DATE? on page 469