Remote control commands
R&S
®
SMA100B
476User Manual 1178.3834.02 ─ 09
Query parameters:
<Board> string
Module name, as queried with the command :
DIAGnostic<hw>:BGINfo:CATalog?.
To retrieve a complete list of all modules, omit the parameter.
The length of the list is variable and depends on the instrument
equipment configuration.
Return values:
<BgInfo> <Module name> <Module stock number incl. variant> <Module
revision> <Module serial number>
List of comma-separated entries, one entry per module.
Each entry for one module consists of four parts that are separa-
ted by space characters.
Example: See Example "Programming example" on page 475.
Usage: Query only
Manual operation: See "Assembly" on page 753
:DIAGnostic:INFO:OTIMe?
Queries the operating hours of the instrument so far.
Return values:
<OperationTime> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example "Programming example" on page 475.
Usage: Query only
Manual operation: See "Operation Time / h" on page 753
:DIAGnostic:INFO:POCount?
Queris how often the instrument has been turned on so far.
Return values:
<PowerOnCount> integer
Range: 0 to INT_MAX
*RST: 0
Example: See Example "Programming example" on page 475.
Usage: Query only
Manual operation: See "Power On Count" on page 753
DIAGnostic subsystem