Analog modulations
R&S
®
SMA100B
104User Manual 1178.3834.02 ─ 09
Double Pulse Width
Sets the width of the second pulse.
Remote command:
[:SOURce<hw>]:PULM:DOUBle:WIDTh on page 585
Pulse Delay
Sets the pulse delay. The pulse delay determines the time that elapses after a trigger
event before pulse modulation starts. The pulse delay is not effective for double pulse
generation.
Remote command:
[:SOURce<hw>]:PULM:DELay on page 584
Double Pulse Delay
Sets the delay from the start of the first pulse to the start of the second pulse.
Remote command:
[:SOURce<hw>]:PULM:DOUBle:DELay on page 584
Pulse Repetition Frequency
Indicates the number of pulses of the repeating signal in a specific time unit.
The parameter is commonly expressed in Hz or in pulses per second (pps).
Remote command:
n.a.
Pulse Output State
Activates the output of the pulse modulation signal.
Remote command:
[:SOURce<hw>]:PGENerator:OUTPut[:STATe] on page 680
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
Pulse Output Polarity
Sets the polarity of the pulse output signal.
Remote command:
[:SOURce<hw>]:PGENerator:OUTPut:POLarity on page 680
Execute Single Trigger
If "Trigger Mode = Single", initiates a single pulse sequence manually.
Remote command:
[:SOURce]:PULM[:INTernal][:TRAin]:TRIGger:IMMediate on page 583
5.4.4.2 Pulse generator > pulse train settings
Option: see Chapter 5.1, "Required options", on page 81.
Modulation settings