Remote control commands
R&S
®
SMA100B
584User Manual 1178.3834.02 ─ 09
Parameters:
<Period> float
The minimum value depends on the installed options
R&S SMAB-K22 or R&S SMAB-K23
Range: 20E-9 to 100
Increment: 5E-9
*RST: 10E-6
Example: See Example "Perform pulse modulation" on page 581.
Manual operation: See "Pulse Period" on page 103
[:SOURce<hw>]:PULM:WIDTh
<Width>
Sets the width of the generated pulse, that means the pulse length. It must be at least
20ns less than the set pulse period.
Parameters:
<Width> float
Range: 20E-9 to 100
Increment: 10E-9
*RST: 2E-6
Example: See Example "Perform pulse modulation" on page 581.
Manual operation: See "Pulse Width" on page 103
[:SOURce<hw>]:PULM:DELay <Delay>
Sets the pulse delay.
Parameters:
<Delay> float
*RST: 1ms
Example: See Example "Perform pulse modulation" on page 581.
Manual operation: See "Pulse Delay" on page 104
[:SOURce<hw>]:PULM:DOUBle:DELay <Delay>
Sets the delay from the start of the first pulse to the start of the second pulse.
Parameters:
<Delay> float
*RST: 1E-6
Example: See Example "Perform pulse modulation" on page 581.
Manual operation: See "Double Pulse Delay" on page 104
SOURce subsystem