Remote control commands
R&S
®
SMA100B
585User Manual 1178.3834.02 ─ 09
[:SOURce<hw>]:PULM:DOUBle:WIDTh <Width>
Sets the width of the second pulse.
Parameters:
<Width> float
Increment: 5E-9
Example: See Example "Perform pulse modulation" on page 581.
Manual operation: See "Double Pulse Width" on page 104
[:SOURce<hw>]:PULM:DOUBle:STATe
<State>
Provided for backward compatibility with former Rohde & Schwarz signal generators.
Works like the command [:SOURce<hw>]:PULM:MODEDOUBle.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
[:SOURce<hw>]:PULM:STATe <State>
Activates pulse modulation.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example: See Example "Perform pulse modulation" on page 581.
Manual operation: See "State" on page 85
[:SOURce<hw>]:PULM:SOURce <Source>
Selects between the internal (pulse generator) or an external pulse signal for the mod-
ulation.
Parameters:
<Source> INTernal | EXTernal
*RST: INTernal
Example: See Example "Perform pulse modulation" on page 581.
Manual operation: See "Source" on page 85
[:SOURce<hw>]:PULM:TTYPe <Source>
Sets the transition mode for the pulse signal.
Parameters:
<Source> SMOothed | FAST
SOURce subsystem