Remote control commands
R&S
®
SMA100B
494User Manual 1178.3834.02 ─ 09
Power and Pulse Data Analysis, Gated Measurements (option R&S SMA100B-
K28)
The power analysis measurement commands are subsumed under the
SENSe[:POWer]:SWEep:… commands. Three measurement modes are available:
Frequency, Power and Time.
The power analysis measurement is started with the
SENSe[:POWer]:SWEep:INITiate command and the measurement result retrieved
with the TRACe[:POWer]:SWEep:… commands.
The four sensors are distinguished by the suffix at the second key word SENSe.
The time gate settings are performed using the CALCulate[:POWer]:SWEep:… com-
mands.
The measurement diagram and results can be stored in a hardcopy with the
SENSe[:POWer]:SWEep:HCOPy:… commands.
General parameter and measurement settings are valid for all connected sensors,
therefore, no suffix is used in these commands.
14.14.1 CALCulate subsystem
:CALCulate[:POWer]:SWEep:FREQuency:MATH<ch>:STATe.............................................494
:CALCulate[:POWer]:SWEep:FREQuency:MATH<ch>:SUBTract........................................ 495
:CALCulate[:POWer]:SWEep:FREQuency:MATH<ch>:XVAL.............................................. 495
:CALCulate[:POWer]:SWEep:FREQuency:MATH<ch>:YVAL.............................................. 495
:CALCulate[:POWer]:SWEep:POWer:MATH<ch>:STATe....................................................496
:CALCulate[:POWer]:SWEep:POWer:MATH<ch>:SUBTract............................................... 496
:CALCulate[:POWer]:SWEep:POWer:MATH<ch>:XVAL..................................................... 497
:CALCulate[:POWer]:SWEep:POWer:MATH<ch>:YVAL..................................................... 497
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:AVERage?.................................................497
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:FEED........................................................498
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:MAXimum?................................................498
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STARt....................................................... 499
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STOP........................................................499
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:STATe....................................................... 499
:CALCulate[:POWer]:SWEep:TIME:MATH<ch>:STATe.......................................................499
:CALCulate[:POWer]:SWEep:TIME:MATH<ch>:SUBTract.................................................. 500
:CALCulate[:POWer]:SWEep:TIME:MATH<ch>:XVAL........................................................500
:CALCulate[:POWer]:SWEep:TIME:MATH<ch>:YVAL........................................................500
:CALCulate[:POWer]:SWEep:FREQuency:MATH<ch>:STATe <State>
Activates the trace mathematics mode for "Frequency" measurement. This feature
enables you to calculate the difference between the measurement values of two
traces. For further calculation, a math result can also be assigned to a trace.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Power sensor measurement subsystems