:CALCulate[:POWer]:SWEep:POWer:MATH<ch>:XVAL <XVal>
Sets the x-axis values for calculating the reference curve in power measurement
mode.
To determine two points ("Point A"/"Point B"), set suffix 1 and 2 in keyword MATH<ch>.
Parameters:
<XVal> float
Range: -145 to 20
Increment: 0.01
*RST: 20
Example:
CALCulate:POWer:SWEep:POWer:MATH1:XVAL 8000
CALCulate:POWer:SWEep:POWer:MATH2:XVAL
1000000000
Sets the power points of the x-axis to calculate the reference
curve.
Options: R&S SMAB-K28
Manual operation: See "Pow (X)" on page 253
:CALCulate[:POWer]:SWEep:POWer:MATH<ch>:YVAL <YVal>
Sets the y-axis values for calculating the reference curve in power measurement
mode.
To determine two points ("Point A"/"Point B"), set suffix 1 and 2 in keyword MATH<ch>.
Parameters:
<YVal> float
Range: -200 to 100
Increment: 0.0001
*RST: -40
Example:
CALCulate:POWer:SWEep:POWer:MATH1:YVAL 20
CALCulate:POWer:SWEep:POWer:MATH2:YVAL -30
Sets the power values on the y-axis for the two values defined
for the x-axis to calculate the reference curve.
Options: R&S SMAB-K28
Manual operation: See "Pow (Y)" on page 254
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:AVERage?
Queries the average power value of the time gated measurement.
Return values:
<Average> float
Range: -1000 to 1000
Increment: 1E-12
*RST: 0