Remote control commands
R&S
®
SMA100B
498User Manual 1178.3834.02 ─ 09
Example:
SENS:SWE:MODE TIME
Activates time mode for power analysis.
CALC:SWE:TIME:GATE:STAT ON
Activates time gated measurement.
SENS:SWE:INIT
Activates a single power analysis measurement.
CALC:SWE:TIME:GATE2:AVER?
Queries the average power in time gate 2 for trace 1 (=default).
Usage: Query only
Options: Option R&S SMAB-K28
Manual operation: See "State - Gate" on page 276
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:FEED
<Feed>
Selects the trace for time gated measurement. Both gates are assigned to the same
trace.
Parameters:
<Feed> TRAC1 | TRAC2 | TRAC3 | TRACe1 | TRACe2 | TRACe3 |
TRAC4 | TRACe4
*RST: TRAC1
Example:
CALC:SWE:TIME:GATE:FEED TRAC2
Assigns the gates to trace 2.
Options: Option R&S SMAB-K28
Manual operation: See "Trace - Gate" on page 275
:CALCulate[:POWer]:SWEep:TIME:GATE<ch>:MAXimum?
Queries the average power value of the time gated measurement.
Return values:
<Maximum> float
Range: -1000 to 1000
Increment: 1E-12
*RST: 0
Example:
SENS:SWE:MODE TIME
Activates time mode for power analysis.
CALC:SWE:TIME:GATE:STAT ON
Activates time gated measurement.
SENS:SWE:INIT
Activates a single power analysis measurement.
CALC:SWE:TIME:GATE2:MAX?
Queries the peak power in time gate 2 for trace 1 (=default).
Usage: Query only
Options: Option R&S SMAB-K28
Power sensor measurement subsystems