Remote control commands
R&S
®
SMA100B
680User Manual 1178.3834.02 ─ 09
[:SOURce<hw>]:PGENerator:OUTPut:POLarity................................................................ 680
[:SOURce<hw>]:PGENerator:OUTPut[:STATe]..................................................................680
[:SOURce<hw>]:PGENerator:STATe................................................................................ 680
[:SOURce<hw>]:PGENerator:OUTPut:POLarity <Polarity>
Sets the polarity of the pulse output signal.
Parameters:
<Polarity> NORMal | INVerted
NORMal
Outputs the pulse signal during the pulse width, that means dur-
ing the high state.
INVerted
Inverts the pulse output signal polarity. The pulse output signal is
suppressed during the pulse width, but provided during the low
state.
*RST: NORMal
Example: See Example "Using pulse generator as source for pulse modu-
lation" on page 679.
Manual operation: See "Pulse Output Polarity" on page 104
[:SOURce<hw>]:PGENerator:OUTPut[:STATe] <State>
Activates the output of the pulse modulation signal.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example: See Example "Using pulse generator as source for pulse modu-
lation" on page 679.
Manual operation: See "Pulse Output State" on page 104
[:SOURce<hw>]:PGENerator:STATe <State>
Enables the output of the video/sync signal.
If the pulse generator is the current modulation source, activating the pulse modulation
automatically activates the signal output and the pulse generator.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example: See Example "Using pulse generator as source for pulse modu-
lation" on page 679.
SOURce subsystem