Remote control commands
R&S
®
SMA100B
657User Manual 1178.3834.02 ─ 09
Example: See Example "Setup an LF sweep" on page 652.
Manual operation: See "Start Frequency/Stop Frequency" on page 194
[:SOURce]:LFOutput<ch>[:STATe] <State>
Activates LF signal output.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example: See Example "Configuring the LF generator" on page 653.
Manual operation: See "State" on page 120
[:SOURce]:LFOutput<ch>:INTernal:VOLTage <Voltage>
Sets the output voltage for the LF generators.
The sum of both values must not exceed the overall output voltage, set with command
[:SOURce]:LFOutput:VOLTage.
Suffix:
<ch>
.
[1]|2
LF1 and LF2
Parameters:
<Voltage> float
Range: 0 to 4
Increment: 0.001
*RST: 1
Example: See Example "Configuring the LF generator" on page 653.
Manual operation: See "Output Voltage" on page 121
[:SOURce]:LFOutput:OFFSet <Offset>
Sets a DC offset at the LF Output.
Parameters:
<Offset> float
Range: depends on lfo voltage
Increment: 0.001
*RST: 0
Example: See Example "Configuring the LF generator" on page 653.
Manual operation: See "DC-Offset" on page 121
SOURce subsystem