Remote control commands
R&S
®
SMA100B
686User Manual 1178.3834.02 ─ 09
[:SOURce<hw>]:POWer:LBEHaviour <Behaviour>
Parameters:
<Behaviour> AUTO | UNINterrupted | MONotone | CVSWr | HDUN
UNINterrupted|MONotone
Uninterrupted level settings and strictly monotone modes.
CVSWr
Constant VSWR
HDUN
High dynamic uninterrupted level settings.
*RST: AUTO
Example:
SOURce1:POWer:LBEHaviour AUTO
Options: R&S SMAB-K724
Manual operation: See "Setting Characteristics" on page 76
[:SOURce<hw>]:POWer:LIMit[:AMPLitude]
<Amplitude>
Limits the maximum RF output level in CW and sweep mode.
It does not influence the "Level" display or the response to the query [:
SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude].
Parameters:
<Amplitude> float
Range: depends on the installed options
Increment: 0.01
*RST: n.a. (factory preset: 30)
Example:
SOURce1:POWer:LIMit:AMPLitude 10
Manual operation: See "Limit" on page 76
[:SOURce<hw>]:POWer:MANual <Manual>
Sets the level for the subsequent sweep step if SWE:POW:MODE MAN.
Use a separate command for each sweep step.
Parameters:
<Manual> float
You can select any level within the setting range, where:
STARt is set with [:SOURce<hw>]:POWer:STARt
STOP is set with [:SOURce<hw>]:POWer:STOP
OFFSet is set with [:SOURce<hw>]:POWer[:LEVel][:
IMMediate]:OFFSet
Range: (STARt + OFFSet) to (STOP + OFFSet)
Increment: 0.01
Default unit: dBm
SOURce subsystem