EasyManua.ls Logo

R&S SMA100B - Page 690

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMA100B
690User Manual 1178.3834.02 ─ 09
EXCLude
Retains the current level when an instrument configuration is
loaded.
*RST: INCLude
Example:
POW:RCL INCL
Takes the level value from an instrument configuration loaded
with command *RCL.
Manual operation: See "Exclude Level" on page 320
[:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude] <Amplitude>
Sets the RF level applied to the DUT.
To activate the RF output use command :OUTPut<hw>[:STATe] ("RF On"/"RF Off").
The following applies POWer = RF output level + OFFSet, where:
POWer is the values set with [:SOURce<hw>]:POWer[:LEVel][:
IMMediate][:AMPLitude]
RF output level is set with [:SOURce<hw>]:POWer:POWer
OFFSet is set with [:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet
Parameters:
<Amplitude> float
The following settings influence the value range:
OFFSet set with the command [:SOURce<hw>]:POWer[:
LEVel][:IMMediate]:OFFSet
Numerical value
Sets the level
UP|DOWN
Varies the level step by step.
The level is increased or decreased by the value set with the
command [:SOURce<hw>]:POWer:STEP[:INCRement].
Range: (Level_min + OFFSet) to (Level_max + OFFStet)
*RST: -30
Default unit: dBm
Example:
POWer -30
Sets the RF level
Example: See also [:SOURce<hw>]:POWer:STEP:MODE on page 688.
Manual operation: See "Amplitude" on page 75
[:SOURce<hw>]:POWer:RANGe:LOWer?
[:SOURce<hw>]:POWer:RANGe:UPPer?
Queries the current interruption-free range of the level.
SOURce subsystem

Table of Contents

Other manuals for R&S SMA100B

Related product manuals