Remote control commands
R&S
®
SMA100B
490User Manual 1178.3834.02 ─ 09
Example:
:OUTPut:ALL:STATe 0
:OUTPut<hw>[:STATe] <State>
Activates the RF output signal.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example:
:OUTPut1 ON
Activates the RF output.
Manual operation: See "RF State/RF ON" on page 65
:OUTPut<hw>[:STATe]:PON <Pon>
Defines the state of the RF output signal when the instrument is switched on.
Parameters:
<Pon> OFF | UNCHanged
*RST: n.a. (factory preset: UNCHanged)
Example:
:OUTPut1:PON OFF
The RF output is deactivated when the instrument is switched
on.
Manual operation: See "Power-On State" on page 345
:OUTPut<hw>:AMODe <AMode>
Sets the step attenuator mode at the RF output.
Parameters:
<AMode> AUTO | FIXed
AUTO
The step attenuator adjusts the level settings automatically,
within the full variation range.
FIXed
The step attenuator and amplifier stages are fixed at the current
position, providing level settings with constant output VSWR.
The resulting variation range is calculated according to the posi-
tion.
*RST: AUTO
Example:
:SOURce1:POWer:ALC:STATe 1
:OUTPut1:AMODe FIXed
Manual operation: See "Mode" on page 215
OUTPut subsystem