Remote control commands
R&S
®
SMA100B
491User Manual 1178.3834.02 ─ 09
:OUTPut<hw>:FILTer:MODE <Mode>
Activates low harmonic filter or enables its automatic switching.
Parameters:
<Mode> ON | AUTO | 1
ON|1
Ensures best low harmonics performance but decreases the
level range
AUTO
Applies an automatically selected harmonic filter that fits to the
current level setting.
*RST: AUTO
Example:
:OUTPut1:FILTer:MODE AUTO
Manual operation: See "Harmonic Filter" on page 77
:OUTPut<hw>:IMPedance?
Queries the impedance of the RF outputs.
Return values:
<Impedance> G1K | G50 | G10K
*RST: G50
Example:
:OUTPut1:IMP?
queries the impedance of RF output.
Response: 50
the impedance is 50 ohms
Usage: Query only
Manual operation: See "RF output impedance" on page 65
:OUTPut:FPRoportional:SCALe <OutpSelScale>
Selects the mode the voltage is supplied depending on the frequency.
The R&S SMA100B supplies the signal at the V/GHz X-Axis connector.
Parameters:
<OutpSelScale> S0V25 | S0V5 | S1V0 | XAXis
S0V25|S0V5|S1V0
Supplies the voltage proportional to the set frequency, derived
from the selected setting.
XAXis
Supplies a voltage range from 0 V to 10 V proportional to the fre-
quency sweep range, set with[:SOURce<hw>]:FREQuency:
STARt and [:SOURce<hw>]:FREQuency:STOP.
*RST: S0V25
OUTPut subsystem