Remote control commands
R&S
®
SMA100B
492User Manual 1178.3834.02 ─ 09
Options: R&S SMAB-B28
Manual operation: See "V/GHz / X-Axis" on page 198
:OUTPut:USER:MARKer <SelUserMarker>
Selects the signal for output at the Marker User1 connector.
Parameters:
<SelUserMarker> MARK | USER
MARK
Assigns a marker signal to the output.
USER
Intended for future use.
*RST: MARK
Options: R&S SMAB-B28
Manual operation: See "Marker / User" on page 199
:OUTPut<hw>:AFIXed:RANGe:LOWer?
:OUTPut<hw>:AFIXed:RANGe:UPPer?
Queries the settable minimum/maximum value in mode :OUTPut:AMODe FIXed, i.e.
when the attenuator is not being adjusted.
See :OUTPut<hw>:AMODe on page 490
Return values:
<Upper> float
Increment: 0.01
Default unit: dBm
Example:
:OUTPut1:AMODe FIXed
:OUTPut1:AFIXed:RANGe:UPPer?
// -27
:OUTPut1:AFIXed:RANGe:LOW?
// -50
Usage: Query only
Manual operation: See "Level Range" on page 215
:OUTPut<hw>:PROTection:CLEar
Resets the protective circuit after it has been tripped.
To define the output state, use the command :OUTPut<hw>[:STATe].
Example:
:OUTPut1:PROT:CLE
Resets the protective circuit of the RF output.
Usage: Event