Remote control commands
R&S
®
SMA100B
647User Manual 1178.3834.02 ─ 09
[:SOURce<hw>]:FREQuency:CENTer <Center>
Sets the center frequency of the sweep.
See Chapter 7.2.1, "Correlating parameters in sweep mode", on page 181.
Parameters:
<Center> float
Range: 300 kHz to RFmax
Increment: 0.01 Hz
*RST: 300E6
Default unit: Hz
Example: See Example "Setup an RF frequency or power sweep"
on page 697
Manual operation: See "Center Frequency" on page 194
[:SOURce<hw>]:FREQuency:SPAN
<Span>
Sets the sapn of the frequency sweep range.
See Chapter 7.2.1, "Correlating parameters in sweep mode", on page 181.
Parameters:
<Span> float
Full freqeuncy range
Increment: 0.01
*RST: 400E6
Example: See Example "Setup an RF frequency or power sweep"
on page 697
Manual operation: See "Span" on page 195
[:SOURce<hw>]:FREQuency:STARt <Start>
Sets the start frequency for the RF sweep.
See Chapter 7.2.1, "Correlating parameters in sweep mode", on page 181.
Parameters:
<Start> float
Range: 300kHz to RFmax
Increment: 0.01Hz
*RST: 100 MHz
Example: See Example "Setup an RF frequency or power sweep"
on page 697
Manual operation: See "Start Frequency/Stop Frequency" on page 194