Remote control commands
R&S
®
SMA100B
689User Manual 1178.3834.02 ─ 09
Note: The command also sets "Variation Step" in the manual control, that means the
user-defined step width for setting the level with the rotary knob or the [Up/Down]
arrow keys.
Parameters:
<Increment> float
Range: 0 to 200
Increment: 0.01
*RST: 1
Default unit: dB
Example: See [:SOURce<hw>]:POWer:STEP:MODE on page 688.
Manual operation: See "Variation Step" on page 73
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:OFFSet <Offset>
Sets the level offset of a downstream instrument.
The level at the RF output is not changed.
To query the resulting level, as it is at the output of the downstream instrument, use the
command [:SOURce<hw>]:POWer[:LEVel][:IMMediate][:AMPLitude].
See "RF frequency and level display with a downstream instrument" on page 66.
Note: The level offset also affects the RF level sweep.
Parameters:
<Offset> float
Range: -100 to 100
Increment: 0.01
*RST: 0
Default unit: dB
Level offset is always expreced in dB; linear units (V, W, etc.) are
not supported
Example:
POWer:OFFSet 10
Sets the RF level offset to 10 dB
Manual operation: See "Offset" on page 76
[:SOURce<hw>]:POWer[:LEVel][:IMMediate]:RCL <Rcl>
Determines whether the current level is retained or if the stored level setting is adopted
when an instrument configuration is loaded.
Parameters:
<Rcl> INCLude | EXCLude
INCLude
Takes the current level when an instrument configuration is loa-
ded.