Remote control commands
R&S
®
SMA100B
643User Manual 1178.3834.02 ─ 09
Example: Setting a phase continuous frequency range
SOURce1:FREQuency:CW 502000000
SOURce1:FREQuency:PHASE:MODE NARR
SOURce1:FREQuency:PHASe:CONTinuous:STATe 1
******************************************
// Query the frequency range
SOURce1:FREQuency:PHASe:CONTinuous:LOW?
// Response: 49370890
SOURce1:FREQuency:PHASe:CONTinuous:HIGh?
// Response: 50421376
******************************************
// Vary the frequency. Within the specified range the signal is phase continuous.
SOURce1:FREQuency:CW 49900000
SOURce1:FREQuency:CW 50000000
SOURce1:FREQuency:CW 50100000
SOURce1:FREQuency:CW 50300000
[:SOURce<hw>]:FREQuency:MODE................................................................................643
[:SOURce<hw>]:FREQuency[:CW|FIXed].........................................................................644
[:SOURce<hw>]:FREQuency[:CW|FIXed]:RCL................................................................. 645
[:SOURce<hw>]:FREQuency:MANual..............................................................................645
[:SOURce<hw>]:FREQuency:MULTiplier.......................................................................... 646
[:SOURce<hw>]:FREQuency:OFFSet.............................................................................. 646
[:SOURce<hw>]:FREQuency:CENTer.............................................................................. 647
[:SOURce<hw>]:FREQuency:SPAN.................................................................................647
[:SOURce<hw>]:FREQuency:STARt................................................................................ 647
[:SOURce<hw>]:FREQuency:STOP.................................................................................648
[:SOURce<hw>]:FREQuency:STEP:MODE...................................................................... 648
[:SOURce<hw>]:FREQuency:STEP[:INCRement]............................................................. 648
[:SOURce<hw>]:FREQuency:PLL:MODE.........................................................................649
[:SOURce<hw>]:FREQuency:PHASe:CONTinuous:HIGH?................................................ 649
[:SOURce<hw>]:FREQuency:PHASe:CONTinuous:LOW?................................................. 649
[:SOURce<hw>]:FREQuency:PHASe:CONTinuous:MODE.................................................650
[:SOURce<hw>]:FREQuency:PHASe:CONTinuous:STATe................................................. 650
[:SOURce<hw>]:FREQuency:MODE <Mode>
Sets the frequency mode for generating the RF output signal. The selected mode
determines the parameters to be used for further frequency settings.
Parameters:
<Mode> CW | FIXed | SWEep | LIST | COMBined
CW|FIXed
Sets the fixed frequency mode. CW and FIXed are synonyms.
The instrument operates at a defined frequency, set with com-
mand [:SOURce<hw>]:FREQuency[:CW|FIXed].