Remote control commands
R&S
®
SMA100B
649User Manual 1178.3834.02 ─ 09
Parameters:
<Increment> float
Range: 0 Hz to RFmax - 100 kHz
Increment: 0.01 Hz
*RST: 1E6
Example: See Example "Frequency configuration" on page 642
Manual operation: See "Variation Step" on page 73
[:SOURce<hw>]:FREQuency:PLL:MODE <Mode>
Selects the PLL (Phase Locked Loop) bandwidth of the main synthesizer.
Parameters:
<Mode> NORMal | NARRow
NORMal
Maximum modulation bandwidth and FM/PhiM deviation.
NARRow
Narrow PLL bandwidth
*RST: NORMal
Example:
SOURce:FREQuency:PLL:MODE NORMal
Manual operation: See "Main PLL Bandwidth" on page 73
[:SOURce<hw>]:FREQuency:PHASe:CONTinuous:HIGH?
Queries the maximum frequency of the frequency range for phase continuous settings.
The maximum frequency of the frequency range depends on the mode selected with
the command [:SOURce<hw>]:FREQuency:PHASe:CONTinuous:MODE.
Return values:
<High> float
Range: 1E5 to 6E9
Increment: 0.01
*RST: 1E9
Default unit: Hz
Example: See Example "Setting a phase continuous frequency range"
on page 643.
Usage: Query only
Manual operation: See "Frequency Range From / To" on page 80
[:SOURce<hw>]:FREQuency:PHASe:CONTinuous:LOW?
Queries the minimum frequency of the frequency range for phase continuous settings.
The minimum frequency of the frequency range depends on the mode selected with
the command [:SOURce<hw>]:FREQuency:PHASe:CONTinuous:MODE.
SOURce subsystem