Remote control commands
R&S
®
SMA100B
650User Manual 1178.3834.02 ─ 09
Return values:
<Low> float
Range: 1E5 to 6E9
Increment: 0.01
*RST: 1E9
Default unit: Hz
Example: See Example "Setting a phase continuous frequency range"
on page 643.
Usage: Query only
Manual operation: See "Frequency Range From / To" on page 80
[:SOURce<hw>]:FREQuency:PHASe:CONTinuous:MODE
<Mode>
Selects the mode that determines the frequency range for the phase continuity.
To query the frequency range, use the commands [:SOURce<hw>]:FREQuency:
PHASe:CONTinuous:HIGH? and [:SOURce<hw>]:FREQuency:PHASe:
CONTinuous:LOW?
Parameters:
<Mode> NARRow | WIDE
NARRow
Small frequency range, asymmetrically around the RF fre-
quency.
WIDE
Large frequency range, symmetrically around the RF frequency.
*RST: NARRow
Example: See [:SOURce<hw>]:FREQuency:PHASe:CONTinuous:
LOW? on page 649
Options: R&S SMAB-B711 (required for frequency range WIDE)
Manual operation: See "Frequency Range" on page 80
[:SOURce<hw>]:FREQuency:PHASe:CONTinuous:STATe <State>
Activates phase continuity of the RF frequency.
The frequency range is limited and varies depending on the set RF frequency. You can
query the range with the commands [:SOURce<hw>]:FREQuency:PHASe:
CONTinuous:HIGH? on page 649 and [:SOURce<hw>]:FREQuency:PHASe:
CONTinuous:LOW? on page 649.
Note: Restricted structure of command line.
In phase continuous mode, the R&S SMA100B only processes the first command of a
command line and ignores further commands if they are on the same line.
SOURce subsystem