Remote control commands
R&S
®
SMA100B
644User Manual 1178.3834.02 ─ 09
SWEep
Sets sweep mode.
The instrument processes frequency (and level) settings in
defined sweep steps.
Set the range and current frequency with the commands:
[:SOURce<hw>]:FREQuency:STARt on page 647 and [:
SOURce<hw>]:FREQuency:STOP on page 648,
[:SOURce<hw>]:FREQuency:CENTer on page 647,
[:SOURce<hw>]:FREQuency:SPAN on page 647,
[:SOURce<hw>]:FREQuency:MANual on page 645
LIST
Sets list mode.
The instrument processes frequency and level settings by
means of values loaded from a list.
To configure list mode settings, use the commands of the Chap-
ter 14.16.7, "SOURce:LIST subsystem", on page 665.
COMBined
Sets the combined RF frequency / level sweep mode.
The instrument processes frequency and level settings in
defined sweep steps.
Set the range and current frequency with the commands:
[:SOURce<hw>]:COMBined:FREQuency:STARt
on page 707 and [:SOURce<hw>]:COMBined:FREQuency:
STOP on page 707,
[:SOURce<hw>]:COMBined:POWer:STARt on page 707
and [:SOURce<hw>]:COMBined:POWer:STOP on page 708
*RST:
CW
Example: See Example "Frequency configuration" on page 642,
and Example "Setup an RF frequency or power sweep"
on page 697
Manual operation: See "State (RF frequency sweep)" on page 188
[:SOURce<hw>]:FREQuency[:CW|FIXed] <Fixed>
Sets the frequency of the RF output signal in the selected path.
The effect depends on the selected mode:
●
In CW mode (FREQ:MODE CW | FIXed), the instrument operates at a fixed fre-
quency.
●
In sweep mode (FREQ:MODE SWE), the value applies to the sweep frequency.
The instrument processes the frequency settings in defined sweep steps.
●
In user mode (FREQ:STEP:MODE USER), you can vary the current frequency step
by step.
Parameters:
<Fixed> float