Remote control commands
R&S
®
SMA100B
707User Manual 1178.3834.02 ─ 09
Manual operation: See "Step Linear/Step Logarithmic" on page 195
[:SOURce<hw>]:COMBined:FREQuency:STARt <CombFreqStart>
Sets the start frequency of the combined RF frequency / level sweep.
See Chapter 7.2.1, "Correlating parameters in sweep mode", on page 181.
Parameters:
<CombFreqStart> float
Range: -59999E5 to 12E9
Increment: 0.01
*RST: 100E6
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
Manual operation: See "Start Frequency/Stop Frequency" on page 194
[:SOURce<hw>]:COMBined:FREQuency:STOP <CombFreqStop>
Sets the end frequency of the combined RF frequency / level sweep.
Parameters:
<CombFreqStop> float
Range: -59999E5 to 12E9
Increment: 0.01
*RST: 500E6
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
Manual operation: See "Start Frequency/Stop Frequency" on page 194
[:SOURce<hw>]:COMBined:POWer:STARt <CombPowStart>
Sets the start level value of the combined RF frequency / level sweep.
See Chapter 7.2.1, "Correlating parameters in sweep mode", on page 181.
Parameters:
<CombPowStart> float
Range: -245 to 120
Increment: 0.01
*RST: -30
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
Manual operation: See "Start Level / Stop Level" on page 196
SOURce subsystem