Remote control commands
R&S
®
SMA100B
706User Manual 1178.3834.02 ─ 09
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
Manual operation: See "Retrace" on page 190
[:SOURce<hw>]:SWEep:POWer:RUNNing?
[:SOURce<hw>]:SWEep[:FREQuency]:RUNNing?
Queries the current sweep state.
Return values:
<State> 1 | ON | 0 | OFF
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
Usage: Query only
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic <Logarithmic>
Sets a logarithmically determined step width for the RF frequency sweep. The value is
added at each sweep step to the current frequency.
See Chapter 7.2.1, "Correlating parameters in sweep mode", on page 181.
Parameters:
<Logarithmic> float
The unit is mandatory.
Range: 0.01 to 100
Increment: 1E-3
*RST: 1
Default unit: PCT
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
Manual operation: See "Step Linear/Step Logarithmic" on page 195
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear] <Linear>
Sets the step width for linear sweeps.
See Chapter 7.2.1, "Correlating parameters in sweep mode", on page 181.
Omit the optional keywords so that the command is SCPI-compliant.
Parameters:
<Linear> float
Range: 0.001 Hz to (STOP - STARt)
Increment: 0.01
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
SOURce subsystem