General instrument functions
R&S
®
SMA100B
374User Manual 1178.3834.02 ─ 09
d) Select "Change Password".
The user password is changed; the user name is displayed ("Security > Password
Management > User Password > User Name").
Changing the default security password
► Note: We recommend that you change the default password before connecting the
instrument to a network.
How to:
● "Changing the default user password of the instrument" on page 373.
● "Changing the default security password" on page 374.
a) Select "System Config > Setup > Security > Security > Password Management
> Security Password".
b) Enter the current password in the "Old Password" field.
The default password is 123456.
c) Enter the new password in the "New Password" and "Confirm Password" fields.
d) Select "Change Password".
Disabling the LAN interface
1. Select "System Config > Setup > Security > Security > LAN Services > Main Ser-
vices".
2. Select "LAN > Off".
All LAN connections and hence all LAN services are disabled.
Disabling LAN services
1. Select "System Config > Setup > Security > Security > LAN Services > Main Ser-
vices".
2. Select, for example, "FTP > Off" or "VNC > Off".
3. Enter the Security Password.
4. Select "Accept".
Activating SMB version 1.0/2.0 client and server
By default, support of SMB client and SMB server of the SMB protocol versions 1.0
and 2.0 is deactivated.
Activate support only, if needed:
1. Select "System Config > Setup > Security > Security > LAN Services > Samba Ser-
vices".
2. Select "SMB 1.0/2.0 Client > On" and "SMB 1.0/2.0 Server > On".
3. Enter the Security Password.
4. Select "Accept".
Using the security settings