Remote control commands
R&S
®
SMA100B
662User Manual 1178.3834.02 ─ 09
Parameters:
<Dwell> float
Range: 0.001 to 100
Increment: 100E-6
*RST: 0.01
Default unit: s
Example: See Example "Setup an LF sweep" on page 652.
Manual operation: See "Dwell Time" on page 192
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:EXECute
Immediately starts an LF sweep.
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:MODE determines which sweep
is executed, e.g. SOURce:LFOutput:SWEep:FREQuency:MODE STEP.
Example:
See Example "Setup an LF sweep" on page 652.
Usage: Event
Manual operation: See "Execute Single Sweep" on page 193
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:MODE <Mode>
Sets the cycle mode of the LF sweep.
Parameters:
<Mode> AUTO | MANual | STEP
AUTO
Performs a complete sweep cycle from the start to the end value
when a trigger event occurs.
The dwell time determines the time period until the signal
switches to the next step.
MANual
Performs a single sweep step when a manual trigger event
occurs.
The trigger system is not active. To trigger each frequency step
of the sweep individually, use the command [:SOURce<hw>]:
LFOutput:FREQuency:MANual on page 655.
STEP
Each trigger command triggers one sweep step only.
The frequency increases by the value set with the coammnds:
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:
LINear] (linear spacing)
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:
LOGarithmic(logarithmic spacing)
*RST:
AUTO
Example: See Example "Setup an LF sweep" on page 652.
SOURce subsystem

Table of Contents