Remote control commands
R&S
®
SMA100B
709User Manual 1178.3834.02 ─ 09
The command triggers one single sweep manually. Therefore, you can use it in manual
sweep mode, selected with the command [:SOURce<hw>]:SWEep:COMBined:MODE
> MANual.
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
Usage: Event
Manual operation: See "Execute Single Sweep" on page 193
[:SOURce<hw>]:SWEep:COMBined:MODE <SweepCombMode>
Sets the cycle mode for the combined frequency / level sweep.
Parameters:
<SweepCombMode> AUTO | MANual | STEP
AUTO
Each trigger event triggers exactly one complete sweep.
MANual
The trigger system is not active. You can trigger every step indi-
vidually by input of the frequencies with the commands [:
SOURce<hw>]:FREQuency:MANual and [:SOURce<hw>]:
POWer:MANual.
STEP
Each trigger event triggers one sweep step.
*RST: AUTO
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
Manual operation: See "Mode" on page 189
[:SOURce<hw>]:SWEep:COMBined:RETRace <RetraceState>
Activates that the signal changes to the start level value while it is waiting for the next
trigger event.
You can enable this feature, when you are working with sawtooth shapes in sweep
mode "Single" or "External Single".
Parameters:
<RetraceState> 1 | ON | 0 | OFF
*RST: 0
Example: See Example "Setup an RF frequency or power sweep"
on page 697.
Manual operation: See "Retrace" on page 190