Remote control commands
R&S
®
SMA100B
639User Manual 1178.3834.02 ─ 09
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 635 .
Usage: Query only
Manual operation: See "User Correction" on page 224
[:SOURce<hw>]:CORRection:ZERoing:STATe <State>
Activates the zeroing procedure before filling the user correction data acquired by a
sensor.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 1
Example: See Example "Fill user correction data with sensor"
on page 636.
Manual operation: See "Fill User Correction Data with Sensor" on page 229
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
Fills the selected user correction list with the level values measured by the power sen-
sor for the given frequencies.
Suffix:
SENSor<ch>
.
Defines the used power sensor, i.e. the sensor whose values are
used.
Example: See Example "Fill user correction data with sensor"
on page 636.
Usage: Event
[:SOURce<hw>]:CORRection[:STATe] <State>
Activates user correction with the currently selected table.
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example: See Example "Create a table with user-defined correction values
for the RF level" on page 635 .
Manual operation: See "State" on page 224
[:SOURce]:CORRection:CSET:CATalog?
Queries a list of available user correction tables.
SOURce subsystem