List and sweep mode
R&S
®
SMA100B
172User Manual 1178.3834.02 ─ 09
Figure 7-2: Schematic representation of a signal generated in list mode (global dwell time)
The list mode is especially useful in high-speed measurements with fast changing
frequency and level settings.
Note that the shown diagrams represent the behaviour in theory. In real signal genera-
tion, the instruments usually have a blank time when the frequency or level changes.
Interactions and characteristics of list and sweep mode
Activating the list mode automatically deactivates all RF and LF sweeps and vice
versa.
In list processing mode, the frequency and level display in the status bar is dis-
abled.
The sweep modes only work with a global dwell time, that means the time intervals
are constant during signal generation.
If you want to hold a running sweep at a specific frequency or level value, enter the
value in the status bar. The sweep stops immediately.
We recommend that you switch off the display update for optimum sweep perfor-
mance, especially with short dwell times
See Chapter 12.1.2, "Display update settings", on page 344.