Improving level performance
R&S
®
SMA100B
269User Manual 1178.3834.02 ─ 09
Duty Cycle ← Notifications - Pulse Data Analysis
Indicates the ratio between the pulse duration (τ) and the pulse period (T) of the mea-
sured pulse signal in per cent:
Duty Cycle = (pulse duration / pulse period) * 100
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DCYCle? on page 540
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DCYCle:DISPlay:
ANNotation[:STATe] on page 541
Pulse Width ← Notifications - Pulse Data Analysis
Indicates the pulse duration of the pulse data measurement in seconds.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DURation? on page 540
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:DURation:DISPlay:
ANNotation[:STATe] on page 541
Pulse Period ← Notifications - Pulse Data Analysis
Indicates the time the pulse signal needs to complete one cycle.
Remote command:
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:PERiod? on page 540
:TRACe<ch>[:POWer]:SWEep:MEASurement:PULSe:PERiod:DISPlay:
ANNotation[:STATe] on page 541
Pulse Off Time ← Notifications - Pulse Data Analysis
Determines the time the pulse signal is low, that means as long as the signal level is
below the proximal value.