Reference oscillator
R&S
®
SMA100B
303User Manual 1178.3834.02 ─ 09
Figure 9-3: Synchronizing instruments with 10
MHz (derived from an external reference frequency)
Ref. Frequency Source = e.g., Rohde & Schwarz signal analyzer
f
Ref
= 10 MHz, 100 MHz, 1 MHz to 100 MHz, 1 GHz** external reference frequency
** = 1 GHz uses Ref In 1 GHz connector
Ref In, Ref Out = Connectors
1 GHz, 1 MHz to 100 MHz at connector Ref In 1 GHz/Ref In and 10 MHz at Ref Out
E.g., external Ref. f
ref
= 100 MHz
Source = "External"
External Reference Frequency = "100 MHz"
Reference Output/1 GHz Reference Output = "10 MHz"
Synchronization Bandwidth = "Narrow"
Using the reference frequency for instruments synchronization