File and data management
R&S
®
SMA100B
312User Manual 1178.3834.02 ─ 09
Overview of the characteristics of the preset functions
Select the preset option that most fits to your particular application:
[Preset]
It is the most frequently used function.
A Preset executes a defined instrument setup to provide an initial instrument state
as a basis for a new configuration. It resets all parameters and switching states,
including also the states of inactive operating modes.
Network, remote access or system settings are retained.
To execute a preset, press the [Preset] key at the front panel.
"Set to Default"
Set To Default relates to individual dialogs or tabs and resets the associated set-
tings of the corresponding dialog. All other settings are retained.
To reset the grouped settings, click "Set To Default".
"Preset this parameter"
Sets a single parameter to its default value.
To reset an individual parameter: Open its context-sensitive menu and select "Pre-
set this parameter...".
"Factory Preset"
A factory preset is the most profound preset function that resets almost all instru-
ment settings, including reference oscillator, network and remote access settings.
Retained are the following settings:
Security, password, and settings protected by these passwords
User-defined data, like setups or data lists
Settings that relate to an integration of the instrument in a measurement setup.
To restore the factory defaults, select System Config > Setup > Settings > Factory
Preset.
Note: Perform a "Factory Preset" only if it is necessary. After a "Factory Preset",
the network connection to the instrument no longer exists.
Restoring the (default) instrument configuration