Remote control commands
R&S
®
SMA100B
545User Manual 1178.3834.02 ─ 09
Example:
TRAC:SWE:MEAS:GATE:DISP:ANN ON
The diagram also shows the gate information.
Options: Option R&S SMAB-K28
Manual operation: See "State - Gate" on page 276
:TRACe[:POWer]:SWEep:MEASurement:MARKer:DISPlay:ANNotation[:STATe]
<State>
Activates the indication of the markers and the marker list in the measurement diagram
and in the hardcopy file.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example:
TRAC:SWE:MEAS:MARK:DISP:ANN ON
The diagram also shows the marker information.
Options: Option R&S SMAB-K28
Manual operation: See "Marker View" on page 277
:TRACe[:POWer]:SWEep:MEASurement:PULSe:DISPlay:ANNotation[:STATe]
<State>
Activates the indication of the pulse data below the measurement diagram and storing
the data in the hardcopy file. The parameters to be indicated can be selected with the
following TRAC:SWE:MEAS:…. commands. Only six parameters are indicated at one
time.
Note: This command is only avalaible in time measurement mode and with R&S
NRPZ81 power sensors.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 1
Example:
TRAC:SWE:MEAS:PULS:DISP:ANN ON
Activates indication of the selected pulse data.
Options: Option R&S SMAB-K28
Manual operation: See "Pulse Data View" on page 278
:TRACe[:POWer]:SWEep:MEASurement:STANdard:DISPlay:ANNotation[:STATe]
<State>
Selects the standard view, i.e. diagram and buttons but no lists are displayed and also
stored in the hardcopy file.
Power sensor measurement subsystems