Remote control commands
R&S
®
SMA100B
565User Manual 1178.3834.02 ─ 09
Parameters:
<State> 1 | ON | 0 | OFF
*RST: 0
Example: See Example "Creating an amplitude modulated RF signal"
on page 562.
Manual operation: See "State" on page 87
[:SOURce<hw>]:AM<ch>:SOURce <Source>
Selects the modulation source for amplitude modulation.
Parameters:
<Source> LF1 | LF2 | NOISe | EXT1 | EXT2 | EXTernal | INTernal
LF1|LF2
Uses an internally generated LF signal.
EXT1|EXT2
Uses an externally supplied LF signal.
NOISe
Uses the internally generated noise signal.
INTernal
Uses the internally generated signal of LF1.
EXTernal
Uses an external LF signal (EXT1).
*RST:
LF1 <AM1>; LF2 <AM2>
Example: See Example "Creating an amplitude modulated RF signal"
on page 562.
Manual operation: See "Source" on page 87
[:SOURce<hw>]:AM<ch>[:DEPTh] <Depth>
Sets the depth of the amplitude modulation in percent.
Suffix:
<ch>
.
1..2
Modulation signal channel
Parameters:
<Depth> float
Range: 0 to 100
Example: See Example "Creating an amplitude modulated RF signal"
on page 562.
[:SOURce<hw>]:AM<ch>:DEPTh:LINear <DepthLin>
Sets the depth of the linear amplitude modulation in percent / volt.