EasyManua.ls Logo

R&S SMA100B - Page 576

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
SMA100B
576User Manual 1178.3834.02 ─ 09
Manual operation: See " Total Deviation" on page 88
[:SOURce<hw>]:FM:RATio <Ratio>
Sets the deviation ratio (path2 to path1) in percent.
Parameters:
<Ratio> float
Range: 0 to 100
Increment: 0.1
*RST: 100
Example: See Example "Creating a frequency modulated RF signal"
on page 572.
Manual operation: See " Ratio Path2/Path1" on page 88
[:SOURce<hw>]:FM:MODE
<Mode>
Selects the mode for the frequency modulation.
Parameters:
<Mode> HBANdwidth | LNOise
HBANdwidth
Selects maximum range for modulation bandwidth.
LNOise
Selects optimized phase noise and spurious characteristics with
reduced modulation bandwidth and FM deviation.
*RST: HBANdwidth
Example: See Example "Creating a frequency modulated RF signal"
on page 572.
Manual operation: See "Mode" on page 89
[:SOURce<hw>]:FM:SENSitivity?
Queries the sensitivity of the externally supplied signal for frequency modulation. The
sensitivity depends on the set modulation deviation.
Return values:
<Sensitivity> float
Sensitivity in Hz/V.
It is assigned to the voltage value for full modulation of the input.
Range: 0 to max
Increment: 0.01
Example: See Example "Using an external signal source" on page 573.
Usage: Query only
SOURce subsystem

Table of Contents

Other manuals for R&S SMA100B