EasyManua.ls Logo

R&S SMA100B - Page 750

R&S SMA100B
822 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting and notifications
R&S
®
SMA100B
750User Manual 1178.3834.02 ─ 09
Selecting the info line also opens the dialog.
The "Static Notifications" dialog lists the recently monitored notifications chro-
nologically and displays additional information on the highlighted message.
The "History" dialog lists the accumulated messages with a short description.
Volatile notifications are reported once. If identical errors occur subsequently,
they are not reported repeatedly.
In the "Info" line, identical errors are displayed repeatedly only if the original
error has already disappeared from the display. If queried by SCPI command,
identical errors are only reported if the original error has already been retrieved
from (and hence not any more present in) the error queue.
Indication and handling of permanent notifications
If any critical error occurs, the R&S SMA100B automatically shows the
icon in the
taskbar. Select the icon to obtain information on the error and the number of occurren-
ces.
The icon is assigned to permanent notifications. The notification and icon are displayed
until the error is eliminated.
Static Notifications/History
Toggles between "Static Notifications" and the "History" tab in the info dialog.
Remote command:
:SYSTem:ERRor[:NEXT]? on page 717
Each time a SYST:ERR:NEXT? query is sent, the eldest entry in the error queue is
returned and at the same time cleared in the list.
:SYSTem:ERRor:STATic? on page 718
Queries the list of all errors.
Clear History
Clears all messages in the "History" tab.
Querying notifications