18 Maintenance and Inspection
18.4 Inspection and Maintenance: O-ring for glass insert
293
GC-2010 Plus
NOTE When you use snoop liquid etc., take care that the liquid is not splashed to the electric wiring and
detector. There is a risk to get an electrical shock.
When using snoop liquid for detecting leaks there is always a possibility that a part   of the liquid
creeps into the pipings which in case of trace analysis can give ghost peaks which may interfere your
peaks of interest. Then it is recommended to use a leak detector instead.
  CAUTION
Do not use the leak detecting fluid nor soapy water for gas leak check
on the connections above the carrier and detector gas controllers
(AFC/APC).
The drips may damage the controller.