Setup
3.4 Inputs and outputs
FST030
Function Manual, 04/2017, A5E35693328-AA
169
3.4.7.5.4 Upper range value
Description
Sets value at 100% of measuring span and corresponds to a loop current of 20 mA.
Setting and Default are sensor specific, depending on the setting in Process value and
depending on the Units setting.
Setting
-3682800.00 kg/h to +3682800.00 kg/h
Default
+3682800.00 kg/h
Related
Path
Setup → Inputs and outputs → Channel 4 - input/output → Current output → Upper range
value
3.4.7.5.5 Lower range value
Description
Sets value at 0% of measuring span and corresponds to a loop current of 0/4 mA.
Setting and Default are sensor specific, depending on the setting in Process value and
depending on the Units setting.
Setting
-10800000000.0 kg/h to +10800000000.0 kg/h
Default
0.00 kg/h
Related
Path
Setup → Inputs and outputs → Channel 4 - input/output → Current output → Lower range
value
3.4.7.5.6 Damping value
Description
Sets the Damping (filtering) to smooth out response to sudden changes in measurement.
An increase to Damping value increases response time of device, and affects digital value
and loop current. If output values are noisy, increase Damping value. For faster response
time, decrease Damping value. Find a value that meets requirements of signal stability and
response time.
Setting
0.0 s to 100.0 s
Default
0.0 s
Related
Path
Setup → Inputs and outputs → Channel 4 - input/output → Current output → Damping value