264 Chapter 16
GPIB Command Reference
ADDRGW
ADDRGW
Syntax ADDRGW <1st>,<2nd>,<3rd>,<4th>
ADDRGW?
Description Sets the gateway IP address when using the 4294A connected to LAN.
NOTE To bring the setting of the changed gateway IP address to take effect, reboot (turn off and
then on again) the 4294A after the setting.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {1st},{2nd},{3rd},{4th}<newline><^END>
Corresponding
key
[Local] - GATEWAY ADDRESS - 1st, 2nd, 3rd, 4th
<1st>
*1
<2nd>
*1
<3rd>
*1
<4th>
*1
Description 1st number of IP
address
2nd number of IP
address
3rd number of IP
address
4th number of IP
address
Range 0 to 255 0 to 255 0 to 255 0 to 255
Initial value 0 0 0 0
Resolution 1 1 1 1
*1.For example, if the IP address is 1.10.100.50, <1st> is 1, <2nd> is 10,
<3rd> is 100, and <4th> is 50.

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals