Chapter 16 365
GPIB Command Reference
OMON
16. GPIB
Command Reference
( O - R )
OMON
Syntax OMON {ON|OFF|1|0}
OMON?
Description Toggles on and off the oscillator level monitor function.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Display] - OSC MON on OFF
OPEP
Syntax OPEP
Description Lists the measurement parameter, sweep type, number of points, delay time (for each
sweep, for each measurement point), sweep range (start value, stop value), and point
averaging count on the LCD screen of the 4294A. (No query)
Corresponding
key
[Copy] - SELECT CONTENTS - OPERATING PARAMETERS 
OSE
Syntax OSE <numeric>
OSE?
Description Sets the value of the Operation Status Enable Register.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
No front panel key is available to execute this function.
Description
ON or 1 Turns on the oscillator level monitor function.
OFF or 0 (initial value) Turns off the oscillator level monitor function.
<numeric>
Description Value of the register
Range 0 to 65535
Initial value 0
Resolution 1

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals