Chapter 16 341
GPIB Command Reference
LISPAN
16. GPIB
Command Reference
( F - N )
LISPAN
Syntax LISPAN {SINGLE|SEGMENT}
LISPAN?
Description Sets the display of the frequency span at list sweep: using the range from the minimum
value to the maximum value in all segments (single span), or using segment-by-segment
allocation.
Parameters
Query response {SINGLE|SEGMENT}<newline><^END>
Corresponding
key
[Sweep] - LIST SPAN [ ]
LISV
Syntax LISV
Description Displays the tabular list of all measurement points and measured values at each
measurement point. Each measured value is updated each time a sweep is completed. To
return to the normal measurement screen, execute the
“RESD” command on page 411. (No
query)
Corresponding
key
[Copy] - SELECT CONTENTS - LIST VALUE
Description
SINGLE
(initial value)
Specifies the single span display as the display method of the
frequency span.
SEGMENT Specifies the segment-by-segment display as the display method of
the frequency span.