360 Chapter 16
GPIB Command Reference
MKRTRMIN
MKRTRMIN
Syntax MKRTRMIN
Description For the active trace (set with the “TRAC” command on page 457), sets the marker position
as the partial search range lower border value. If the marker function is off, executing this
command causes an error and the command is ignored. (No query)
Corresponding
key
[Search] -SEARCH RANGE MENU - MKR → MINIMUM
MKRVAL?
Syntax MKRVAL?
Description For the active trace (set with the “TRAC” command on page 457), reads out the marker
position measurement parameter value.
Query response {numeric}<newline><^END>
If the measurement parameter is a scalar value (for other than COMPLEX Z-Y), the
measurement parameter value is read out. If the measurement parameter is a vector value
(for COMPLEX Z-Y), the readout is as follows:
• If the polar chart format is selected, the amplitude is read out.
• If the complex plane format is selected, resistance (for trace A) or conductance (for
trace B) is read out.
Corresponding
key
No front panel key is available to execute this function.