ACCUD
16. GPIB
Command Reference
( * , A - E )
The Agilent 4294A commands
This section describes the GPIB commands specific to the 4294A.
ACCUD
Syntax ACCUD {ON|OFF|1|0}
ACCUD?
Description Enables/disables the display mode to accumulate traces in which they are not cleared at
each sweep.
Parameters
Query response {1|0}<newline><^END>
Corresponding
key
[Display] - ACCUMULATE on OFF
ADDRCONT
Syntax ADDRCONT <numeric>
ADDRCONT?
Description Sets the GPIB address of the controller of the 4294A.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Corresponding
key
[Local] - ADDRESS: CONTROLLER
Description
ON or 1 Enables the trace accumulating display.
OFF or 0
(Initial value)
Disables the trace accumulating display (the trace is cleared at each
sweep and only the latest trace is displayed).
<numeric>
Description GPIB address of the controller
Range 1 to 30
Initial value 21
Resolution 1

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals