Chapter 16 385
GPIB Command Reference
OUTPMEMO?
16. GPIB
Command Reference
( O - R )
OUTPMEMO?
Syntax OUTPMEMO?
Description Reads out the values (complex number) of all measurement points in the memory array
(refer to
“Internal data arrays” on page 81). (Query only)
Query response {numeric 1},{numeric 2},..,{numeric NOP×2-1},{numeric NOP×2}<newline><^END>
Reads out the real part and imaginary part of each measurement point value (complex
number) separately. Where, NOP is the number of points, and n is an integer between 1 and
NOP.
{numeric n×2-1}: Measurement parameter value of the n-th measurement point (real
part).
{numeric n×2}: Measurement parameter value of the n-th measurement point
(imaginary part).
Corresponding
key
No front panel key is available to execute this function.
OUTPMEMOP?
Syntax OUTPMEMOP? <numeric 1>
Description Reads out the value (complex number) of the specified measurement point in the memory
array (refer to
“Internal data arrays” on page 81). (Query only)
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric 2},{numeric 3}<newline><^END>
The real part and imaginary part are read out separately as shown below.
{numeric 2}: Measurement parameter value of the specified measurement point
(real part).
{numeric 3}: Measurement parameter value of the specified measurement point
(imaginary part).
Corresponding
key
No front panel key is available to execute this function.
<numeric 1>
Description Number of the measurement point you want to read out (the
number assigned in order from the left edge assuming that the
number of the leftmost measurement point is 1)
Range 1 to the number of points
Resolution 1

Table of Contents

Other manuals for Agilent Technologies 4294A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4294A and is the answer not in the manual?

Agilent Technologies 4294A Specifications

General IconGeneral
BrandAgilent Technologies
Model4294A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals